Abstract
Background
The current investigation examined the inter- and intra-tester reliability of knee joint angle measurements using a flexible Penny and Giles Biometric® electrogoniometer. The clinical utility of electrogoniometry was also addressed.
Methods
The first study examined the inter- and intra-tester reliability of measurements of knee joint angles in supine, sitting and standing in 35 healthy adults. The second study evaluated inter-tester and intra-tester reliability of knee joint angle measurements in standing and after walking 10 metres in 20 healthy adults, using an enhanced measurement protocol with a more detailed electrogoniometer attachment procedure. Both inter-tester reliability studies involved two testers.
Results
In the first study, inter-tester reliability (ICC[2,10]) ranged from 0.58-0.71 in supine, 0.68-0.79 in sitting and 0.57-0.80 in standing. The standard error of measurement between testers was less than 3.55° and the limits of agreement ranged from -12.51° to 12.21°. Reliability coefficients for intra-tester reliability (ICC[3,10]) ranged from 0.75-0.76 in supine, 0.86-0.87 in sitting and 0.87-0.88 in standing. The standard error of measurement for repeated measures by the same tester was less than 1.7° and the limits of agreement ranged from -8.13° to 7.90°. The second study showed that using a more detailed electrogoniometer attachment protocol reduced the error of measurement between testers to 0.5°.
Conclusion
Using a standardised protocol, reliable measures of knee joint angles can be gained in standing, supine and sitting by using a flexible goniometer. © 2008 Piriyaprasarth et al; licensee BioMed Central Ltd.
The current investigation examined the inter- and intra-tester reliability of knee joint angle measurements using a flexible Penny and Giles Biometric® electrogoniometer. The clinical utility of electrogoniometry was also addressed.
Methods
The first study examined the inter- and intra-tester reliability of measurements of knee joint angles in supine, sitting and standing in 35 healthy adults. The second study evaluated inter-tester and intra-tester reliability of knee joint angle measurements in standing and after walking 10 metres in 20 healthy adults, using an enhanced measurement protocol with a more detailed electrogoniometer attachment procedure. Both inter-tester reliability studies involved two testers.
Results
In the first study, inter-tester reliability (ICC[2,10]) ranged from 0.58-0.71 in supine, 0.68-0.79 in sitting and 0.57-0.80 in standing. The standard error of measurement between testers was less than 3.55° and the limits of agreement ranged from -12.51° to 12.21°. Reliability coefficients for intra-tester reliability (ICC[3,10]) ranged from 0.75-0.76 in supine, 0.86-0.87 in sitting and 0.87-0.88 in standing. The standard error of measurement for repeated measures by the same tester was less than 1.7° and the limits of agreement ranged from -8.13° to 7.90°. The second study showed that using a more detailed electrogoniometer attachment protocol reduced the error of measurement between testers to 0.5°.
Conclusion
Using a standardised protocol, reliable measures of knee joint angles can be gained in standing, supine and sitting by using a flexible goniometer. © 2008 Piriyaprasarth et al; licensee BioMed Central Ltd.
| Original language | English |
|---|---|
| Article number | 6 |
| Pages (from-to) | 1-10 |
| Number of pages | 10 |
| Journal | BMC Musculoskeletal Disorders |
| Volume | 9 |
| DOIs | |
| Publication status | Published - 22 Jan 2008 |
| Externally published | Yes |