Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines

Wayne A. Hing, Keith Rome, Alyse F. M. Cameron

Research output: Contribution to journalArticleResearchpeer-review

16 Citations (Scopus)

Abstract

Background: Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HDII Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.

Methods: The abductor hallucis muscle of both the left and right feet of thirty asymptomatic participants was imaged and then measured using both ultrasound machines. Interclass correlation coefficients (ICC) with 95% confidence intervals (CI) were used to calculate both within and between session intra-tester reliability. Standard error of the measurement (SEM) calculations were undertaken to assess difference between the actual measured score across trials and the smallest real difference (SRD) was calculated from the SEM to indicate the degree of change that would exceed the expected trial to trial variability.

Results: The ICCs, SEM and SRD for dorso-plantar thickness and medial-lateral length were shown to have excellent to high within and between-session reliability for both ultrasound machines. The between-session reliability indices for cross-sectional area were acceptable for both ultrasound machines.

Conclusion: The results of the current study suggest that regardless of the type ultrasound machine, intra-tester reliability for the measurement the abductor hallucis muscle parameters is very high.

Original languageEnglish
Article number33
Number of pages8
JournalJournal of Foot and Ankle Research
Volume2
DOIs
Publication statusPublished - 2009
Externally publishedYes

Cite this

@article{03eea15209284e77b84808c35b254bec,
title = "Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines",
abstract = "Background: Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HDII Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.Methods: The abductor hallucis muscle of both the left and right feet of thirty asymptomatic participants was imaged and then measured using both ultrasound machines. Interclass correlation coefficients (ICC) with 95{\%} confidence intervals (CI) were used to calculate both within and between session intra-tester reliability. Standard error of the measurement (SEM) calculations were undertaken to assess difference between the actual measured score across trials and the smallest real difference (SRD) was calculated from the SEM to indicate the degree of change that would exceed the expected trial to trial variability.Results: The ICCs, SEM and SRD for dorso-plantar thickness and medial-lateral length were shown to have excellent to high within and between-session reliability for both ultrasound machines. The between-session reliability indices for cross-sectional area were acceptable for both ultrasound machines.Conclusion: The results of the current study suggest that regardless of the type ultrasound machine, intra-tester reliability for the measurement the abductor hallucis muscle parameters is very high.",
author = "Hing, {Wayne A.} and Keith Rome and Cameron, {Alyse F. M.}",
year = "2009",
doi = "10.1186/1757-1146-2-33",
language = "English",
volume = "2",
journal = "Journal of Foot and Ankle Research",
issn = "1757-1146",
publisher = "BioMed Central Ltd.",

}

Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines. / Hing, Wayne A.; Rome, Keith; Cameron, Alyse F. M.

In: Journal of Foot and Ankle Research, Vol. 2, 33, 2009.

Research output: Contribution to journalArticleResearchpeer-review

TY - JOUR

T1 - Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines

AU - Hing, Wayne A.

AU - Rome, Keith

AU - Cameron, Alyse F. M.

PY - 2009

Y1 - 2009

N2 - Background: Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HDII Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.Methods: The abductor hallucis muscle of both the left and right feet of thirty asymptomatic participants was imaged and then measured using both ultrasound machines. Interclass correlation coefficients (ICC) with 95% confidence intervals (CI) were used to calculate both within and between session intra-tester reliability. Standard error of the measurement (SEM) calculations were undertaken to assess difference between the actual measured score across trials and the smallest real difference (SRD) was calculated from the SEM to indicate the degree of change that would exceed the expected trial to trial variability.Results: The ICCs, SEM and SRD for dorso-plantar thickness and medial-lateral length were shown to have excellent to high within and between-session reliability for both ultrasound machines. The between-session reliability indices for cross-sectional area were acceptable for both ultrasound machines.Conclusion: The results of the current study suggest that regardless of the type ultrasound machine, intra-tester reliability for the measurement the abductor hallucis muscle parameters is very high.

AB - Background: Diagnostic ultrasound provides a method of analysing soft tissue structures of the musculoskeletal system effectively and reliably. The aim of this study was to evaluate within and between session reliability of measuring muscle dorso-plantar thickness, medio-lateral length and cross-sectional area, of the abductor hallucis muscle using two different ultrasound machines, a higher end Philips HDII Ultrasound machine and clinically orientated Chison 8300 Deluxe Digital Portable Ultrasound System.Methods: The abductor hallucis muscle of both the left and right feet of thirty asymptomatic participants was imaged and then measured using both ultrasound machines. Interclass correlation coefficients (ICC) with 95% confidence intervals (CI) were used to calculate both within and between session intra-tester reliability. Standard error of the measurement (SEM) calculations were undertaken to assess difference between the actual measured score across trials and the smallest real difference (SRD) was calculated from the SEM to indicate the degree of change that would exceed the expected trial to trial variability.Results: The ICCs, SEM and SRD for dorso-plantar thickness and medial-lateral length were shown to have excellent to high within and between-session reliability for both ultrasound machines. The between-session reliability indices for cross-sectional area were acceptable for both ultrasound machines.Conclusion: The results of the current study suggest that regardless of the type ultrasound machine, intra-tester reliability for the measurement the abductor hallucis muscle parameters is very high.

U2 - 10.1186/1757-1146-2-33

DO - 10.1186/1757-1146-2-33

M3 - Article

VL - 2

JO - Journal of Foot and Ankle Research

JF - Journal of Foot and Ankle Research

SN - 1757-1146

M1 - 33

ER -