Reliability and concurrent validity of postural asymmetry measurement in adolescent idiopathic scoliosis

Ashleigh Prowse, Berit Aslaksen, Marie Kierkegaard, James Furness, Paul Gerdhem, Allan Abbott

Research output: Contribution to journalArticleResearchpeer-review

11 Citations (Scopus)
201 Downloads (Pure)

Abstract

AIM: 

To investigate the reliability and concurrent validity of the Baseline(®) Body Level/Scoliosis meter for adolescent idiopathic scoliosis postural assessment in three anatomical planes.

METHODS: 

This is an observational reliability and concurrent validity study of adolescent referrals to the Orthopaedic department for scoliosis screening at Karolinska University Hospital, Stockholm, Sweden between March-May 2012. A total of 31 adolescents with idiopathic scoliosis (13.6 ± 0.6 years old) of mild-moderate curvatures (25° ± 12°) were consecutively recruited. Measurement of cervical, thoracic and lumbar curvatures, pelvic and shoulder tilt, and axial thoracic rotation (ATR) were performed by two trained physiotherapists in one day. The intraclass correlation coefficient (ICC) was used to determine the inter-examiner reliability (ICC2,1) and the intra-rater reliability (ICC3,3) of the Baseline(®) Body Level/Scoliosis meter. Spearman's correlation analyses were used to estimate concurrent validity between the Baseline(®) Body Level/Scoliosis meter and Gold Standard Cobb angles from radiographs and the Orthopaedic Systems Inc. Scoliometer.

RESULTS: 

There was excellent reliability between examiners for thoracic kyphosis (ICC2,1 = 0.94), ATR (ICC2,1 = 0.92) and lumbar lordosis (ICC2,1 = 0.79). There was adequate reliability between examiners for cervical lordosis (ICC2,1 = 0.51), however poor reliability for pelvic and shoulder tilt. Both devices were reproducible in the measurement of ATR when repeated by one examiner (ICC3,3 0.98-1.00). The device had a good correlation with the Scoliometer (rho = 0.78). When compared with Cobb angle from radiographs, there was a moderate correlation for ATR (rho = 0.627).

CONCLUSION: 

The Baseline(®) Body Level/Scoliosis meter provides reliable transverse and sagittal cervical, thoracic and lumbar measurements and valid transverse plan measurements of mild-moderate scoliosis deformity.

Original languageEnglish
Pages (from-to)68-76
Number of pages9
JournalWorld Journal of Orthopedics
Volume8
Issue number1
DOIs
Publication statusPublished - 18 Jan 2017