High-resolution metal artifact reduction MR imaging of the lumbosacral plexus in patients with metallic implants

Shivani Ahlawat, Steven E. Stern, Allan J. Belzberg, Jan Fritz

Research output: Contribution to journalArticleResearchpeer-review

7 Citations (Scopus)

Abstract

Objective: To assess the quality and accuracy of metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) for the diagnosis of lumbosacral neuropathies in patients with metallic implants in the pelvis. 

Materials and methods: Twenty-two subjects with lumbosacral neuropathy following pelvic instrumentation underwent 1.5-T MARS MRI including optimized axial intermediate-weighted and STIR turbo spin echo sequences extending from L5 to the ischial tuberosity. Two readers graded the visibility of the lumbosacral trunk, sciatic, femoral, lateral femoral cutaneous, and obturator nerves and the nerve signal intensity of nerve, architecture, caliber, course, continuity, and skeletal muscle denervation. Clinical examination and electrodiagnostic studies were used as the standard of reference. Descriptive, agreement, and diagnostic performance statistics were applied. 

Results: Lumbosacral plexus visibility on MARS MRI was good (4) or very good (3) in 92% of cases with 81% exact agreement and a Kendall’s W coefficient of 0.811. The obturator nerve at the obturator foramen and the sciatic nerve posterior to the acetabulum had the lowest visibility, with good or very good ratings in only 61% and 77% of cases respectively. The reader agreement for nerve abnormalities on MARS MRI was excellent, ranging from 95.5 to 100%. MARS MRI achieved a sensitivity of 86%, specificity of 67%, positive predictive value of 95%, and negative predictive value of 40%, and accuracy of 83% for the detection of neuropathy. 

Conclusion: MARS MRI yields high image quality and diagnostic accuracy for the assessment of lumbosacral neuropathies in patients with metallic implants of the pelvis and hips.

Original languageEnglish
Pages (from-to)897-908
Number of pages12
JournalSkeletal Radiology
Volume46
Issue number7
DOIs
Publication statusPublished - Jul 2017

Fingerprint

Lumbosacral Plexus
Artifacts
Metals
Magnetic Resonance Imaging
Obturator Nerve
Pelvis
Muscle Denervation
Femoral Nerve
Acetabulum
Sciatic Nerve
Thigh
Hip
Skeletal Muscle
Sensitivity and Specificity
Skin

Cite this

@article{30f041067f984845bfacec5330729e4e,
title = "High-resolution metal artifact reduction MR imaging of the lumbosacral plexus in patients with metallic implants",
abstract = "Objective: To assess the quality and accuracy of metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) for the diagnosis of lumbosacral neuropathies in patients with metallic implants in the pelvis. Materials and methods: Twenty-two subjects with lumbosacral neuropathy following pelvic instrumentation underwent 1.5-T MARS MRI including optimized axial intermediate-weighted and STIR turbo spin echo sequences extending from L5 to the ischial tuberosity. Two readers graded the visibility of the lumbosacral trunk, sciatic, femoral, lateral femoral cutaneous, and obturator nerves and the nerve signal intensity of nerve, architecture, caliber, course, continuity, and skeletal muscle denervation. Clinical examination and electrodiagnostic studies were used as the standard of reference. Descriptive, agreement, and diagnostic performance statistics were applied. Results: Lumbosacral plexus visibility on MARS MRI was good (4) or very good (3) in 92{\%} of cases with 81{\%} exact agreement and a Kendall’s W coefficient of 0.811. The obturator nerve at the obturator foramen and the sciatic nerve posterior to the acetabulum had the lowest visibility, with good or very good ratings in only 61{\%} and 77{\%} of cases respectively. The reader agreement for nerve abnormalities on MARS MRI was excellent, ranging from 95.5 to 100{\%}. MARS MRI achieved a sensitivity of 86{\%}, specificity of 67{\%}, positive predictive value of 95{\%}, and negative predictive value of 40{\%}, and accuracy of 83{\%} for the detection of neuropathy. Conclusion: MARS MRI yields high image quality and diagnostic accuracy for the assessment of lumbosacral neuropathies in patients with metallic implants of the pelvis and hips.",
author = "Shivani Ahlawat and Stern, {Steven E.} and Belzberg, {Allan J.} and Jan Fritz",
year = "2017",
month = "7",
doi = "10.1007/s00256-017-2630-9",
language = "English",
volume = "46",
pages = "897--908",
journal = "Skeletal Radiology",
issn = "0364-2348",
publisher = "Springer",
number = "7",

}

High-resolution metal artifact reduction MR imaging of the lumbosacral plexus in patients with metallic implants. / Ahlawat, Shivani; Stern, Steven E.; Belzberg, Allan J.; Fritz, Jan.

In: Skeletal Radiology, Vol. 46, No. 7, 07.2017, p. 897-908.

Research output: Contribution to journalArticleResearchpeer-review

TY - JOUR

T1 - High-resolution metal artifact reduction MR imaging of the lumbosacral plexus in patients with metallic implants

AU - Ahlawat, Shivani

AU - Stern, Steven E.

AU - Belzberg, Allan J.

AU - Fritz, Jan

PY - 2017/7

Y1 - 2017/7

N2 - Objective: To assess the quality and accuracy of metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) for the diagnosis of lumbosacral neuropathies in patients with metallic implants in the pelvis. Materials and methods: Twenty-two subjects with lumbosacral neuropathy following pelvic instrumentation underwent 1.5-T MARS MRI including optimized axial intermediate-weighted and STIR turbo spin echo sequences extending from L5 to the ischial tuberosity. Two readers graded the visibility of the lumbosacral trunk, sciatic, femoral, lateral femoral cutaneous, and obturator nerves and the nerve signal intensity of nerve, architecture, caliber, course, continuity, and skeletal muscle denervation. Clinical examination and electrodiagnostic studies were used as the standard of reference. Descriptive, agreement, and diagnostic performance statistics were applied. Results: Lumbosacral plexus visibility on MARS MRI was good (4) or very good (3) in 92% of cases with 81% exact agreement and a Kendall’s W coefficient of 0.811. The obturator nerve at the obturator foramen and the sciatic nerve posterior to the acetabulum had the lowest visibility, with good or very good ratings in only 61% and 77% of cases respectively. The reader agreement for nerve abnormalities on MARS MRI was excellent, ranging from 95.5 to 100%. MARS MRI achieved a sensitivity of 86%, specificity of 67%, positive predictive value of 95%, and negative predictive value of 40%, and accuracy of 83% for the detection of neuropathy. Conclusion: MARS MRI yields high image quality and diagnostic accuracy for the assessment of lumbosacral neuropathies in patients with metallic implants of the pelvis and hips.

AB - Objective: To assess the quality and accuracy of metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) for the diagnosis of lumbosacral neuropathies in patients with metallic implants in the pelvis. Materials and methods: Twenty-two subjects with lumbosacral neuropathy following pelvic instrumentation underwent 1.5-T MARS MRI including optimized axial intermediate-weighted and STIR turbo spin echo sequences extending from L5 to the ischial tuberosity. Two readers graded the visibility of the lumbosacral trunk, sciatic, femoral, lateral femoral cutaneous, and obturator nerves and the nerve signal intensity of nerve, architecture, caliber, course, continuity, and skeletal muscle denervation. Clinical examination and electrodiagnostic studies were used as the standard of reference. Descriptive, agreement, and diagnostic performance statistics were applied. Results: Lumbosacral plexus visibility on MARS MRI was good (4) or very good (3) in 92% of cases with 81% exact agreement and a Kendall’s W coefficient of 0.811. The obturator nerve at the obturator foramen and the sciatic nerve posterior to the acetabulum had the lowest visibility, with good or very good ratings in only 61% and 77% of cases respectively. The reader agreement for nerve abnormalities on MARS MRI was excellent, ranging from 95.5 to 100%. MARS MRI achieved a sensitivity of 86%, specificity of 67%, positive predictive value of 95%, and negative predictive value of 40%, and accuracy of 83% for the detection of neuropathy. Conclusion: MARS MRI yields high image quality and diagnostic accuracy for the assessment of lumbosacral neuropathies in patients with metallic implants of the pelvis and hips.

UR - http://www.scopus.com/inward/record.url?scp=85016438701&partnerID=8YFLogxK

U2 - 10.1007/s00256-017-2630-9

DO - 10.1007/s00256-017-2630-9

M3 - Article

VL - 46

SP - 897

EP - 908

JO - Skeletal Radiology

JF - Skeletal Radiology

SN - 0364-2348

IS - 7

ER -